What HPLC purity and mass-spectrometry identity each establish, what a chromatogram shows, where both methods stop being able to tell you anything, and the criteria a lot is released against.
These are the two questions every analytical certificate answers, and they are answered by different instruments because they are not the same question.
A material can be extremely pure and still be the wrong compound. It can also be unambiguously the right compound and be half of what is in the vial. Neither figure substitutes for the other, and a certificate that reports only one of them has told you half of what you need.
High-performance liquid chromatography pushes a dissolved sample through a packed column under pressure. Components travel at different speeds depending on how strongly they interact with the column packing, so a mixture that entered as one band leaves as a series of separated bands. A detector at the far end records each one as it emerges.
Reverse-phase is the common configuration for peptides: a non-polar stationary phase, a polar mobile phase, and a gradient that becomes progressively less polar over the run so that increasingly hydrophobic components are released in turn. Detection is typically ultraviolet absorbance in the region where the peptide bond itself absorbs.
Almost always it means area percent: the area under the main peak, divided by the total area under every integrated peak in the run, expressed as a percentage. It is a statement about proportion within what the method detected.
That last clause is the whole caveat. Area percent assumes every component responds to the detector in proportion to how much of it is present. Components that absorb weakly at the detection wavelength are under-represented. Components that do not absorb at all — many salts, residual solvents, water — are not in the denominator, because the detector never saw them. A peptide reported at 98.6% by RP-HPLC can still be a minority of the vial by mass.
This is why net peptide content, water content and counter-ion content are separate determinations, and why a purity figure alone does not tell you how much peptide is in a vial.
Mass spectrometry ionises the sample, accelerates the ions through a field, and measures their mass-to-charge ratio. For a peptide of known sequence the theoretical mass is a calculation, so the measurement is a direct comparison: the observed mass either matches the expected mass within the method's tolerance, or it does not.
Electrospray ionisation typically produces a series of multiply-charged ions rather than one peak, and the neutral mass is reconstructed from that series. This is normal, and a spectrum showing a charge-state envelope is not a defect.
Coupling liquid chromatography to a mass spectrometer separates the sample first and identifies each separated component second. That answers a question neither instrument answers alone: not just how large the impurity peaks are, but what they are. A truncated sequence, a deletion, an oxidised residue and a completely unrelated contaminant are all "an impurity peak" to an ultraviolet detector, and are distinguishable by mass.
Mass is not sequence. Two peptides containing the same residues in a different order have the same mass, and isomeric substitutions can be invisible to a mass measurement alone. Establishing sequence requires fragmentation — tandem mass spectrometry — or an orthogonal method such as Edman degradation. A certificate reporting "identity confirmed by MS" is reporting a mass match, and a mass match is strong evidence, not proof of sequence.
A chromatogram plots detector response against time. Time runs left to right; height is how much signal the detector saw at that moment. Each peak is a component leaving the column.
Stated plainly, because a testing programme that only publishes what its methods can show is more useful than one that implies they show everything.
A lot moves through a fixed sequence of states, and a state is only advanced by a person. Nothing promotes itself, and a lot that fails a check does not proceed to the next one.
There are two honest outcomes. Either an assignable laboratory error is documented and the assay is repeated, with both the original and the repeat retained in the record — or there is no assignable error, the result stands, and the lot is not released. Repeating an assay until it produces an acceptable number is the practice this policy exists to prevent.
A published lot record is permanent and is never deleted. A withdrawn lot keeps its page and states that it was withdrawn. Corrections are appended as amendments, never written over the original. A supplier who can delete the record of a bad lot has not published evidence; they have published marketing.
A supplier testing its own material and reporting its own result is asking the buyer to trust the supplier twice. Third-party testing replaces one of those with a laboratory that has no interest in the outcome, and it only means anything if the laboratory is named — an unnamed "independent lab" is an unverifiable assertion.
A supplier-issued certificate is not worthless. It is simply a different class of evidence, and it should be labelled as what it is rather than presented as independent testing.